CHAPTER 6
Operation
6-20
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
E
U
1. Rinse a diluent dispensing container (mayer, beaker, etc.)
with CELLPACK to remove dust, etc.
2. Dispense CELLPACK into the diluent dispensing
container.
3. Use a pipette (160
µ
L) to dispense 160
µ
L of CELLPACK
into a microtube.
4. Use a pipette (40
µ
L) to dispense 40
µ
L of blood into a
microtube.
5. Cap and mix the microtube well.
Sample analysis
1. Make sure that the instrument is in READY status.
The READY LED should be lit.
2. Press the MANUAL key on the Main Unit panel keypad.
The Sample No. Setting screen will appear on the LCD.
3. Using the numeric keys, input the sample ID number. Or,
read the barcode using a handheld barcode scanner
(option).
4. When analysis mode is not set in capillary mode, press
the
key and change the set item to “Mode”. Then using
the
and
keys to set the capillary mode.
5. Check the discrete setting, and if necessary, press
key
to change the set item to “Discrete”. Then using the
and
keys, set the discrete.
6. After all the settings are completed, press the ENTER key
to confirm.
Information
HPC analysis and body fluid analysis are not possible
in capillary mode.
READY LED
N e x t N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 6
1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
1 2 3 4 5
C B C C B C C B C C B C C B C
D I F F D I F F
N R B C N R B C N R B C
R E T R E T
S a m p l e 1 : N o r m a l 2 : H P C 3 : B o d y F l u i d
D P N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
C D N R
N u m
D P
X m
C a p i l l a r y
N o t R e a d y
< S e l e c t M o d e a n d N o . >
S a m p l e N o .
M o d e
D i s c r e t e
M a n u a l C a p i l l a r y C l o s e d
1 2 3
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...