5. Mixing Motor
This program performs test operation of the reaction chamber
mixing motor.
Execute this test when a “Mixing Motor Error” occurs and
analysis cannot be performed. If the test is normally
completed, analysis will be possible after the system enters
the Ready status.
1. From the Main Menu screen on the Main Unit, select
“Test” on the function menu.
The Test Menu screen will appear.
2. Select “Motor” on the function menu of the Test Menu
screen.
The Motor Test screen will appear.
3. Select “5. Mixing Motor” by moving the cursor with
or
key on the Motor Test screen.
4. Select “Execute” on the function menu of the Motor Test
screen. The reaction chamber mixing motor will start
running, and the motor revolution speed will be displayed
on the Testing screen.
Normal range: 1300 < Revolution speed < 2000
5. Approximately 30 seconds later, the reaction chamber
mixing motor will stop rotating, and the test will be
completed.
6. To cancel the motor test, select “Cancel” on the function
menu, or press the RETURN key.
SRV
This program performs test operation of the sample rotor
valve (SRV) and sets the valve in the correct initial position.
Execute this test when a “SRV Lower Position ERR” or “SRV
Upper Position ERR” occurs and analysis cannot be
performed. If the test is normally completed, analysis will be
possible after the system enters the Ready status.
1. Open the front cover of the Main Unit.
2. From the Main Menu screen on the Main Unit, select
“Test” on the function menu.
The Test Menu screen will appear.
3. Select “SRV” on the function menu of the Test Menu
screen, then select “Execute”.
The sample rotor valve will start operation and perform
volume determination. Then the valve will return to the
initial position.
4. To cancel the SRV test, select “Cancel” on the function
menu, or press the RETURN key.
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...