CHAPTER 7
Quality control
7-18
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
Variable target function and erroneous data
In the X/L-J control, there are restrictions on the import of
control data into the control chart to ensure the smooth
operation of the variable target function. The specific
restriction is that control data beyond three times the range of
the control range set in advance must not be imported.
Assume that the control limits are the values given above,
with the implementation of X/L-J control. If the control data is
5.21 million, the message “Check QC Chart” will be displayed,
but the data can be entered on the chart. If the control data is
5.61 million however, it should not be entered on the QC
Chart, and the message “Reanalyze sample” will be
displayed.
Setting variable target
1. Click a parameter you wish to set in the list.
2. Click the “Variable Target” button.
The selected parameters are set for variable target. The
target column of the list is left blank for parameters set for
variable target .
Information
• When “Variable Target” is selected for quality
control, set the limits manually or read the assay
limits from the CD-ROM.
• When “C New” is selected in the Lot column,
the target and limit range cannot be changed or set
using the “Target/Limit” button (it is inactive).
Select a lot you wish to change or set: “Current” or
“New”.
UL
RBC
LL
520
500
480
The upper limits to import the control data
560
440
The upper limits for the control limit (UL)
The control target values (TARGET)
The lower limits for the control limit (LL)
The lower limits to import the control data
LIMIT
Example:
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...