CHAPTER 9
Cleaning and Maintenance
9-48
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
E
U
Adjusting the pressure to 0.16 MPa
This pressure is used to supply sheath reagent to the detector
block.
1. Display the Sensor 1 screen on the LCD screen.
See “Chapter 9: 9.21 Displaying the pressure or vacuum”.
2. Pull out the knob for the 0.16 MPa regulator to release the
lock.
3. Adjust the pressure while watching the pressure displayed
on the Sensor 1 screen.
To increase the pressure, turn the knob clockwise.
Normal range: 0.16 ± 0.001 MPa
4. Check that the pressure displayed on the Sensor 1 screen
is the predetermined pressure.
5. If the displayed pressure does not match the
predetermined pressure, adjust it by repeating the
operations 3 and 4.
6. When the adjustment is completed, press the adjustment
knob for the 0.16 MPa regulator to lock.
Adjusting the pressure to 0.07 MPa
This pressure is used to discharge waste fluid and mix the
sample.
1. Display the Sensor 1 screen on the LCD screen.
See “Chapter 9: 9.21 Displaying the pressure or vacuum”.
2. Pull out the knob for the 0.07 MPa regulator to release the
lock.
3. Adjust the pressure while watching the pressure displayed
on the Sensor 1 screen.
To increase the pressure, turn the knob clockwise.
Normal range: 0.07 ± 0.001 MPa
Information
Always adjust pressure so as to increase to the
predetermined level. If the pressure is too high, lower
it below that level once; then raise it to the set level.
Low
High
Adjustment
knob
Low
High
Adjustment
knob
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...