CHAPTER 7
Quality control
Sysmex XE-5000 Instructions for Use
7-7
R
e
vi
se
d A
p
ri
l 20
07
7.5
QC data display
After the analysis is completed, the results are displayed.
If X control is used as control data, the data in the “Mean”
column is stored; in the L-J control, the latest data is stored.
The control data will be stored when it is displayed on the
LCD screen.
X control display
L-J control display
If the message “CHECK CONTROL CHART” appears in the
action message column, the analyzed data has exceeded the
control limit. If the message “REANALYZE CONTROL”
appears, the analyzed data is more than 3 times the limit.
The “+” and “-” signs in the judgment column indicate which
parameters are outside the control limits and which are above
or below the limits. (When analyzed data is more than 3 times
the limit, “+” and “-” will be highlighted.)
N e x t N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 6
D P N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
M a n u a l
N o r m a l
N o t R e a d y
X 1
X 2
M e a n
J u d g e
W B C
R B C
H G B
H C T
M C V
M C H
M C H C
7 . 7 1
4 . 5 4
1 2 . 3
3 2 . 7
7 7 . 1
2 5 . 7
3 3 . 3
7 . 7 3
4 . 5 5
1 2 . 3
3 2 . 9
7 7 . 3
2 5 . 9
3 3 . 7
7 . 7 2
4 . 5 4
1 2 . 3
3 2 . 8
7 7 . 2
2 5 . 8
3 3 . 5
C D N R
L O T : 1 2 3 4 5 6 7 8 < E x e c u t e X b a r >
N u m
D P
X m
O K
C a n c e l
N e x t N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 6
D P N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
D 1
J u d g e
M a n u a l
N o r m a l
N o t R e a d y
W B C
R B C
H G B
H C T
M C V
M C H
M C H C
7 . 7 1
4 . 5 4
1 2 . 3
3 2 . 7
7 7 . 1
2 5 . 7
3 3 . 3
C D N R
L O T : 1 2 3 4 5 6 7 8 < E x e c u t e L - J >
N u m
D P
X m
O K
C a n c e l
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...