2.11 Marking on the instrument
Main Unit - Front View
(1)
(2)
(3)
(3)
(1)
(8)
(9)
(6)
(2)
(5)
(7)
(4)
*1
(Inside)
WARNING
Be sure to set the stop bar when opening
to avoid injury.
WARNING
Never touch the detector when the power
of the Main Unit is turned ON. Otherwise,
electrical shock may result.
CAUTION
1. Dye solution replacement method
• For SNR, always replace dye solution and lyse reagent simultaneously.
• For RED, always replace dye solution and diluent simultaneously.
• For FFS, the message “Replace FFS” is displayed after the analysis of
2000 cycles. When doing so, always press Reagent or Confirm after
replacing with new FFS.
2. When the message “Execute Rinse Flowcell” is displayed, always perform
Maintenance – 4. Rinse Flowcell.
3. To clean the IMI detector aperture, first run Maintenance – 1. Drain IMI,
then turn off the main unit power.