CHAPTER 6
Operation
Sysmex XE-5000 Instructions for Use
6-13
R
e
vi
se
d A
p
ri
l 20
07
• Manual closed mode
In manual closed mode, the sampler is used to aspirate the
sample, without opening the cap of the sample tube. This
mode is basically the same as manual, HPC and body fluid
modes; mixing and continuous analysis can not be
performed automatically.
6.16 Preparing sample analysis
Samples
• Collect the specified amount of blood (corresponding to the
amount of EDTA anticoagulant).
• Using CELLPACK dispensed ahead of time, dilute the
sample to a 1:5 ratio for capillary mode.
• Sample volumes
Manual, capillary and manual closed mode
HPC analysis and Body fluid analysis
Sampler mode
Note:
HPC analysis and body fluid analysis can only be
performed in manual mode.
Analysis
mode
Manual
Capillary
Manual Closed
Required
Sample
1.0 mL or more
40
µ
L or more
1.0 mL or more
Aspirated
Sample
Approx. 130
µ
L
Approx. 130
µ
L
Approx. 200
µ
L
Analysis mode
HPC
Body fluid
Required Sample
1.0 mL or more
500
µ
L or more
Aspirated Sample
Approx. 130
µ
L
Approx. 130
µ
L
Diameter of Sample Tube
12 mm
15 mm
Required Sample
1.0 to 5 mL
1.0 to 7 mL
Aspirated Sample
200
µ
L
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...