STROMATOLYSER-IM (SIM) replacement procedure
1. Prepare a new reagent and confirm that its expiration date
has not expired.
2. Open the seal on the new reagent container.
3. After opening the seal, check that there is an O-ring on
the tip of the joint.
4. Pull the collar of the one-touch coupling up, and remove it
from the empty reagent container.
Caution!
• When replacing STROMATOLYSER-IM, always wear
gloves.
• Use reagent that has been left at room temperature
(15–30°C) for at least 24 hours.
• In handling a reagent that may have frozen, follow
the precautions given on the package.
Otherwise, analysis may not be performed correctly.
• Make sure there is no dust etc. sticking to the joint
between the reagent container and the one-touch
coupling.Use alcohol (isopropanol 70%) to clean any
away.
Otherwise, analysis may not be performed correctly.
• When changing the reagent container, take care not
to pinch or kink the tube.
Otherwise, analysis may not be performed correctly.
O
-
ring
Collar
One-touch
coupling
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...