15
1.4.5
Electrical
WARNING
Failure to comply with the following may result in serious or potentially fatal
bodily injuries to the operator or those in the area.
Only properly trained and qualified personnel should be permitted access to
any internal parts of the x-ray system. Live electrical terminals may be
deadly; be sure line disconnect switches are opened and other appropriate
precautions are taken before opening access doors, removing enclosure
panels, or attaching accessories.
Do not remove the flexible high voltage cables from the x-ray tube housing or
high voltage transformer or the access covers from the generator until the
main and auxiliary power supplies have been disconnected.
When disconnecting high voltage cables, they must be grounded
immediately in order to dissipate any electrical charge that may remain on
the cables or the tube.
1.4.6 Mechanical
WARNING
Particular care should be taken when servicing the inside of the tubestand. There
is an extreme threat of mechanical pinching between the vertical slide and
counterweight due to their close proximity and opposite directions of motion.
All of the movable assemblies and parts of x-ray equipment should be operated
with care. Only properly trained and qualified personnel should be permitted
access to any internal parts of the x-ray system.