2.0 MECHANICAL
ASSEMBLY
2.1 Radiographic
Table
1.
During shipment, the tabletop is banded to the table base. Cut loose the
banding and set the table top aside for assembly later. Loosen the hold-
down screws underneath the front rail of the table to release the grid cabinet.
Remove packing material from rear of grid cabinet.
2.
Install the five leveling feet by threading them into the holes at the two front
corners of the table and at the ends and middle of the back rail. The two feet
with a 4" long body should be installed at the front corners of the table. The
three feet with a 1" long body should be installed at the back rail. (In the
accessory bag there are six blue and six red “U” shaped plastic spacers.
When leveling the table fully tighten the foot until the shoulder of the foot is
firmly against either these spacers or the bottom of the table/back rail). Use
of the spacers makes the table base/leveling feet union a solid “one piece”
structure, minimizing any wobble or table base motion.
Position the table close to its final location, using the leveling feet and
spacers to bring the table close to level at this time. Final leveling can be
done at end of assembly.
2.2 High Voltage Generator and Transformer placement
1.
Verify that the oil level in the HV transformer is approximately 1/2" to 3/4"
below the underside of the cover. Refill with "Shell Diala oil AX" or remove
oil as necessary.
2.
Using a piece of cardboard to protect the floor, push the transformer onto the
shelf in the table by tilting the transformer and sliding it up onto the right side
of the shelf. Once the transformer is in place, loosen the screw on the vent
plug to allow for oil expansion. Cover the vent plug and flange with one of
the loose fitting plastic covers provided with the HV transformer to prevent
contaminants from entering the oil.
3.
Place the Power Module close to the front left side of the table but not inside,
leaving access on all sides. This allows the installer to establish the proper
wiring, routing and lengths for all cables.