10.0 TROUBLESHOOTING
10.1 Overview
The 20 kHz HF APR generator has an extensive internal monitoring system. A number of
error codes are displayed to alert the operator or service engineer to the status of the
system. Some error codes are momentary alerts, and will automatically reset once the
offending button has been released. Other error codes indicate simple or serious
problems with the system. The chart below shows all error codes, what they mean, and
how to correct them.
Table 10-1: Table of Error Codes
Error Code
Likely cause
How to clear the
fault
Secondary
actions
INVALID
kVp LIMIT
Operator
Selected value
out of range
Release kVp button
Unit resets to
valid techniques
INVALID
mAs LIMIT
Operator
Selected value
out of range
Release mAs
button
Unit resets to
valid techniques
INVALID
TUBE LIMIT
Operator
Selected value
out of range
Release the
button
kVp and mAs are
reset
INVALID
kW LIMIT
Operator
Selected value
out of range
Release the
button
Unit resets to
valid techniques
INVALID
TIME LIMIT
Operator
Selected value
> 5 seconds
Release the
button
Unit resets to
valid
techniques
INVALID
TIME LIMIT
Operator
Selected value
< .01 seconds
Release the
button
Unit resets to
valid techniques