InnoVet Select
™
HF APR
Service Manual
50
7.3.3 Power Supply Test and Fuse Designation
Verify the following output voltages from the Power Supplies referenced to ground.
Table 7-3: Fusing for DC Voltages
F3
0.5 Amp, Fast Acting
+5 VDC
F4
0.5 Amp, Time Delay
+12 VDC
F5
0.5 Amp, Time Delay
-12 VDC
The following fuses are also present.
Table 7-4: Fusing for AC Voltages
Fuse #
Rating
Function
F1
50 Amp, Time Delay
Mains
F2
50 Amp, Time Delay,
Mains
F6
10 Amp, Time Delay
240 VAC autotransformer return
F7
10 Amp, Time Delay
240 VAC autotransformer hot
F8
7 Amp, Time Delay
Collimator supply
F9
5 Amp, Time Delay
Electric lock supply (24 VDC)
F10
1 Amp, Time Delay
Relay pcb supply
F11
0.8 Amp, Time Delay
12 VAC Console supply
F12
1.5 Amp, Time Delay
Bucky supply (Not used for vet)
F13
8 Amp, Time Delay
240 VAC rotor supply
F14
8 Amp, Time Delay
240 VAC rotor common
F15
3 Amp, Time Delay
240 VAC power supply pcb input