Service Manual
52
8.2 TUBE ROLL CAL Screen and Calibration
This selection will allow the installer to calibrate the tube roll display. Ensure the
“roll” potentiometer has been pre-set before the calibration procedure. (See
section 2.4 for “roll” pot pre-set procedure)
Once in the CAL BOOTUP screen, select TUBE ROLL CAL. The following screen will be
displayed.
B A C K
Tube Roll Calibration
1.
Rotate the entire operator’s console and tube head to the left until the primary beam
is horizontal. This will be the +90 degree point. Use a level to verify this position.
Push the
SAVE +90 DEG
up/down buttons to input this setting. Unit will give an
audible tone and display
SAVING
. +90 degree calibration is complete.
2.
Rotate the entire operator’s console and tube head until the primary beam is
pointed vertically downward. This will be the 0 degree point. Use a level to verify
this position. Push the
SAVE 0 DEG
up/down buttons to input this setting. Unit will
give an audible tone and display
SAVING
. 0 degree calibration is complete.
3.
Rotate the entire operator’s console and tube head to the right until the primary
beam is horizontal. This will be the –90 degree point. Use a level to verify this
position. Push the
SAVE -90 DEG
up/down buttons to input this setting. Unit will
give an audible tone and display
SAVING
. -90 degree calibration is complete.
4.
Tube roll calibration is complete. Verify various tube angle displays and repeat
calibration if necessary.
NOTE: The “roll” angle will only be displayed in normal operating RAD mode. The
“roll” display will not be displayed in any of the set-up utility menus or during an
error condition in normal RAD mode.
TUBE ROLL
CAL
SAVE +90 DEG
SAVE 0 DEG
SAVE –90 DEG
LOAD DEFAULT