InnoVet Select
™
HF APR
Service Manual
67
Error Code
Likely cause
How to clear the fault
Secondary
actions
ERROR
mA PRESENT
RESTART
SYSTEM.
Check 01003-
000at H9 pins 5
and 6.
Cycle power. Find
defective
components.
ERROR
kV PRESENT
RESTART
SYSTEM
Check 01003-
000at H9 pins 1
and 3.
Cycle power. Find
defective
components.
ERROR
LOW FIL AMPS
Open secondary,
filament pcb, or
poor mA leading
edge calibration.
Press any key.
Check mA
output
with/without
feedback.
ERROR
HI FIL AMPS
Filament pcb, or
poor mA leading
edge calibration.
Press any key.
Check filament
pcb, and mA
output with and
without
feedback.
ERROR
EXP. RELEASE
Prep/exp switch
or operator error.
Press any key.
ERROR
IPM OVERLOAD
RESTART
SYSTEM
Excessive current
through inverter.
Cycle power. Find
defective
components.
See "Trouble-
shooting
Guide" below.
ERROR
kVp OVERLOAD
Secondary arcing,
kVp > 135.
Press any button.
Inspect
secondary and
calibration.
ERROR kV
IMBALANCE
Anode/Cathode
kV feedback
missing or wrong
polarity
Cycle power, fix kV
feedback circuit
see "trouble
codes" below
ERROR
mA OVERLOAD
RESTART
SYSTEM
Secondary arcing,
calibration, mA is
> 380.
Cycle power.
Inspect
secondary and
calibration.