InnoVet Select
™
HF APR
Service Manual
40
6.7 Field Installation of a Line Matching Transformer
Generally, if a line-matching transformer is required it is installed in the field. In the
event that the supply voltage is below 230 VAC and a transformer is not installed,
follow these instructions. If a line-matching transformer is required, contact the
factory to order a line matching transformer kit, part number
L246
.
Warning
:
Make absolutely certain that the AC mains supply is off and that
the DC capacitor bank voltage within the generator is fully
discharged before beginning this procedure. The capacitor bank
voltage requires a minimum of 20 minutes to discharge. Verify
that no voltages are present.
1. Turn off the incoming power supply at the service disconnect. Remove the Power
Module cover (see the service manual for details as required).
2. At the bottom of the Power Module locate two #5/16 threaded holes. Place mounting
bracket above them and secure with 5/16-18 x ½ lg bolts and locking washers.
3. Carefully place the toroid line-matching transformer with rubber gasket inside the
bracket (as shown below) so as not to scratch the transformer insulation on the
bracket. Orient the transformer so that the leads exit the bracket to the top and right.
4. Secure the transformer to the bracket using #3/8-16 x 4-1/2” long hex bolt and locking
hardware. The bolt is to be snug, but not tightened so excessively that it crushes the
epoxy center of the toroid style transformer.
X-FMR
RUBBER PAD
MOUNTING
BRACKET
BOLT, 5/16-18 X 1/2
& 5/16 LOCKING
WASHER
FLAT WASHER
LOCK WASHER
BOLT, 3/8-16 X 4"
LONG
WARNING:
DO NOT
CONNECT ANY WIRES TO
THIS MOUNTING BOLT