InnoVet Select
™
HF APR
Service Manual
30
3.0 FINAL MECHANICAL ADJUSTMENTS
3.1 Leveling The Tubearm
An adjustment screw for leveling the tube arm is provided on the front of tubestand,
just below the tube arm base. First remove the decorative hole cover to access the
adjustment screw. Next turn the screw clockwise to raise the tube arm or
counterclockwise to lower the tube arm. Replace the decorative hole cover when
finished leveling.
3.2 40 Inch SID Detents
To position the tube 40" from either the table top or the film cassette, detents are
provided. These detents are factory adjusted, however some adjustment is
provided if needed. Access to the adjustments is from the rear of the tubestand
.
3.3 Verification Of X-Ray To Light Field Coincidence
The coincidence of x-ray to light field must be verified. Specific instructions can be
found in the manual
for the G800 collimator.