InnoVet Select
™
HF APR
Service Manual
49
Table 7-2: DIP2 Functions (APR Console PCBA)
Dip Switch
OPEN (OFF)
CLOSED (ON)
DIP2-1
Normal Operation
Not used
DIP2-2
Normal Operation
PCMCIA card write protect
DIP2-3
Normal Operation
Simulator mode (demo)
DIP2-4
Normal Operation
Simulator mode (demo)
DIP2-5
Normal Operation
Simulator mode (demo)
DIP2-6
Normal Operation
Simulator mode (demo)
DIP2-7
Latching Prep Disabled
Latching Prep Enabled
DIP2-8
Normal Operation
Not used
7.3 Power-Up
7.3.1 Start-Up
Turn the main disconnect switch ON. Press the Operator Console ON key and observe
the display screen during the start-up phase. The screen will display the date and time,
and the choice of APR or MANUAL (2 Point) mode of operation.
Note that in MANUAL mode of operation, EXPOSURE COUNT can be selected to
momentarily display the total number of exposures taken on this x-ray control.
7.3.2 Status LEDs
Upon power up, there are a number of LEDs, which will illuminate on boards within the
power module. Verify that the following are true.
01003-000System Controller PCB: LED1 lights will strobe back and forth across the bar.
K650 Power Supply PCB: LED 1, 2 and 3
01710-000 Power on/off PCB: LED 1, 2 and 3
K660 Relay PCB: LED 6
01770-000Filament PCB: LED 1 and 2, LED 3 for Large filament select