InnoVet Select
™
HF APR
Service Manual
51
8.0 SYSTEM CONFIGURATION USING THE OPERATOR CONSOLE
Note: Throughout this manual "CAPS" text is used to indicate the wording on
the console or display.
8.1 CAL BOOTUP screen.
This screen is the entry point for all Operator Console
based system configuration such as tube type and image receptor configuration. It is also
where manipulation of the anatomical technique database and calibration screens are
accessed.
To access the CAL BOOTUP screen: Turn the control OFF. Press and hold the
BACK button while pressing the ON button.
The CAL BOOTUP screen will display as
shown below.
BACK
DATE AND TIME: Allows the user to program local time.
TUBE ROLL CAL: Allows the installer to calibrate the tube roll potentiometer
XRAY CAL:
Installers access to configuration and calibration screens.
EDIT APR MODE: Allows the user to make changes to the technique database.
RENAME TECH:
Allows the user to add more techniques/change view names.
MEMORY FUNC: Allows the installer to save configuration and/or technique
data to a manufacturer supplied memory card and transfer it between
sites.
EXIT TO RAD:
Returns the operator to APR or MANUAL modes of operation.
SET DATE CAL BOOTUP
& TIME
EDIT APR MODE
REMANE
TECH
TUBE ROLL CAL MEMORY FUNC
XRAY CAL
EXIT TO RAD