12.2 Accuracy of kVp and mAs Outputs
The published specifications for this generator are ±(5%+1) kVp, and ±(10%+1) mAs.
Local regulations may be more or less stringent. Make exposures, recording the selected
kVp and mAs below. Calculate the required system accuracy by subtracting test
equipment error using the methods described above and record the error percentage.
Table 12-1: Table for Recording kVp and mAs Outputs
Selected kVp
Measured kVp
Selected mAs
Measured mAs
kVp error %
mAs error %
NOTE: If the unit does not meet the accuracy specifications, view the mA and
kVp waveforms, verify calibration, and check for line voltages below 230 VAC in
either idle or full load exposure mode. Correct the problem and retest the
system.