6.8 High
Voltage
Cables
Connect the high voltage cables from the HV Transformer to the x-ray tube. With
the vent to the front, the cathode (-) receptacle is on the left, near the red and black
mAs meter jumpers. The anode (+) receptacle is on the right. Verify polarity and
that all connections are made correctly.
CAUTION: The high voltage cable terminal pins are delicate. Take particular
care to handle them carefully.
NOTE:
The HV Cable knurled nuts are delicate and can be damaged by
dropping or banging on the floor. This can deform the threads,
making it difficult to install them fully into the HV receptacles
Use a clean, dry, lint-free cloth and alcohol to clean the insulating surfaces of the
terminals and receptacles. Use particular care in cleaning the flat insulating
surface. Do not touch the insulating surface after cleaning. Apply vapor-proofing
compound to both cable plugs at each end. See the manufactures high voltage
cable installation manual for specific cable preparation.
The high voltage cables will route into the table base as shown in the figure
below (top view looking down with tabletop panel removed).
Figure 11 - High Voltage Cable Routing
Top view looking down w/tabletop panel removed