5.0 GENERATOR
PRE-INSTALLATION
5.1 Introduction
This section describes procedures necessary for the installation of the following
units:
•
Table front cover
•
High Voltage Transformer
•
Power
Module
•
Connections to ancillary devices common to an X-ray room
This material is presented in sequence to facilitate an orderly and safe installation.
Complete calibration procedures are presented in Section 8.
5.2 Installation Items and Test Equipment
The following test equipment is required for the installation:
•
Fluke Digital Multimeter or equivalent
•
Machlett Dynalyzer or equivalent high voltage divider for kVp and mA values
•
Keithley dosimeter or equivalent for linearity and reproducibility
•
Tektronics 210 dual trace storage oscilloscope or equivalent for waveform
shape
The following miscellaneous items are required or may prove useful for the
installation:
•
Silicone Insulating Grease - DC4 vapor-proofing compound, or equivalent.
•
Alcohol
•
Standard tool kit
•
Victoreen mAs meter or equivalent
•
Keithley Non-invasive kVp meter or equivalent