InnoVet Select
™
HF APR
Service Manual
65
Error Code
Likely cause
How to clear the fault
Secondary
actions
INVALID
BUTTON
Press another
button to make
desired change.
Release the
button
Unit resets to
valid
techniques
INVALID
mA LIMIT
User attempted to
select < 50 mA or
> 300 mA.
Release the
button
Unit resets to
valid
techniques
INVALID
cm LIMIT
User attempted to
select a cm value
beyond APR data.
Release the
button
Unit resets to
valid
techniques
ERROR
MICRO-P
RESTART
SYSTEM
Electrical noise,
HV arcing, +5v
supply, 01003-
000pcb.
Cycle power.
The filament is
disabled.
ERROR
NO COMM.
RESTART
SYSTEM
Communication
cables
Cycle power, find
break in serial link.
ERROR
+/- 12 v LIMIT
RESTART
SYSTEM
Power supply pcb,
lock noise, or DC
fusing
Cycle power, put
diodes across locks,
find damaged
component.
ERROR
EEP R/W
The calibration
data (checked
during operation)
is corrupted.
Cycle power. Call for
service. See
"troubleshooting
guide" below.
ERROR
EEP
CHECKSUM
Calibration data
(checked at power
up) is corrupted.
Press any button.
See "troubleshooting
guide" below.
ERROR
PREP STUCK
RESTART
SYSTEM
Prep signal or
switch may be
shorted.
Find the damaged
component.
ERROR
EXPOSE STUCK
RESTART
SYSTEM
Expose signal or
switch may be
shorted.
Find the damaged
component.