11.5 The CARD TO CONS screen
The CARD TO CONSOLE screen allows APR data from the Memory Card to be
downloaded into the Operator Console. This can be useful if this site is using the
same film/screen combination as a previous installation which has modified APR
techniques, or if the x-ray service company has an established technique database
they like to offer. A confirmation screen will ask if you wish to copy all APR
techniques from the selected memory card with a NO or YES response. After the
data transfer is complete, turn the console off and remove the card.
NOTE: If EEPROM CHECKSUM ERROR is diplayed, return to the
MEMORY FUNCTIONS screen and select PM TO CONS. This will re-copy the
existing calibration data from the power module into the operator console and clear
the error.
11.6 The CONS TO CARD screen
The CONSOLE TO CARD screen allows uploading APR technique database
information from an Operator Console into the Memory Card. This can be useful for
on-site back-up storage of a modified APR database, or for carrying a modified
database to be downloaded at another site. A confirmation screen will ask if you
wish to copy all APR techniques from the Operator Console to the memory card
with a NO or YES response. The modified techniques will remain in the Operator
Console, but now there will be a back-up copy in the Memory Card.
WARNING: The memory card can only hold one full set of APR
data. Saving data from the console to the card overwrites data
which was previously stroed.