92
CHAPTER 6
Press the
CYCLE
button until
SET: AZT RANGE
is displayed.
Press the
ENTER
button to select the
SET: AZT RANGE
option.
Press either the
UP
or
DOWN ARROW
buttons to adjust the number to
"0"
.
Example:
Press the
ENTER
button to store the AZT selected and return to the
SET: AZT RANGE
Menu option.
Press the
PWR / MENU
button twice to go directly to the normal operating mode.
Note:
Pressing the
PWR / MENU
button before pressing the
ENTER
button will cancel that
configuration
change.
Note:
If the
PWR / MENU
button is held too long the meter will display all 8’s and the meter will
turn off.
Disconnect one load cell from the VSL Vulcoder of the channel to be tested.
Plug that one lead into the connector on the Check-Out box. Make sure the load switch is pointing
away from the red dot.
Note:
The Check-Out box is needed to properly zero the electronics. If the
Check-Out box is not used then you may receive incorrect readings.
The meter will display the load cell preload number. The preload number should not exceed
+12,000 lb to -12,000 lb.
If the preload number is 12,000 lb to -12,000 lb, remove the lead from the Check-Out box and
reconnect it to the load cell. Test the next load cell in the same manner. If the preload number is
within specification for all load cells, reconnect and secure all load cell leads. Reset the current Cal
Number then reset the current Tare weight.
Note:
If the preload number is out of specification, clean
the load cell connector and cable connector with cotton swabs and isopropyl alcohol and dry
thoroughly with a hair dryer and
DO NOT OVERHEAT
the connector.
Check the electrical isolation of the load cell by performing the Leakage Test, (Section 6.2.1).
June 21, 2006
VULCAN ON-BOARD SCALES 1-800-237-0022
Doc. 44-10032-001 Rev. C
©
STRESS-TEK, INC. 2006