background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for ADM Series

Page 1: ......

Page 2: ...CONTENTS CAUTIONS 1 INSTALLATION 2 SPECIFICATIONS 3 KEYBOARD LAYOUT AND DESCRIPTION 4 INITIAL SETUP 5 INSTRUCTION FOR USE 6 TROUBLE SHOOTING 7 DAILY CARE AND MAINTENANCE ...

Page 3: ... where or when the environment condition falls beyond as those listed on SPECIFICATIONS 3 Do not attempt to open this unit or conduct any trouble shootings other than those listed on TROUBLE SHOOTING 4 Do not submerge scale into water 5 Do not use pressurized water nozzle for cleaning 6 Do not use this scale for chemical substances or in any environment that contains chemical vapor 7 Precision mea...

Page 4: ...ber cap FIG 1 C 1 3 Place the stainless steel platter FIG 1 D onto the plastic platter 1 4 Power on the scale a Before plugging the power adaptor into electricity outlet check and make sure that the input voltage of power adaptor matches with the output voltage of the outlet If not do not plug in and contact your dealer immediately b Before first time use plug the adaptor into the wall outlet and ...

Page 5: ...FIG 1 FIG 2 ...

Page 6: ...ive Display 5 x 24mm LCD display Operation Environment 0o 40o C 32o 104o F Non condensed R H 85 Power source By Built in rechargeable battery or power adaptor Specifications subject to change without notice The minimum load of 15kg for Non OIML application is 25e NOTE 1 OIML Legal for trade 1 3000 Non OIML Not legal for trade 1 6000 1 7500 NOTE 2 Display resolution of avoirdupois weight units lb l...

Page 7: ... or reset displayed weight to zero when the scale is not loaded 2 TARE KEY Press this key to tare off the weight of a container during normal operation 3 MODE KEY Press this key to shift among different weight units and display parameters 4 OFF KEY Press this key to turn scale off ...

Page 8: ...arge the scale immediately Failure to do so may cause unrecoverable damage to the rechargeable battery inside scale 9 IN CHARGE INDICATOR Red color Recharging battery Green color Charging completed 10 lb lb UNIT INDICATOR This indicator appears to indicate the weighing unit being selected Two avoirdupois weight units are available and this indicator means the weighing unit is at decimal pound e g ...

Page 9: ...to see if the display and backlight if purchased are functioning properly a Scale is off b Press and hold TARE then press ON ZERO c Scale displays F 1 d Press TARE once e Scale displays F 2 f Press MODE and all segments will light up to allow user to check the condition of display g Press TARE for next function setup or ON ZERO to restart the scale 4 2 2 Select Auto Power Off Setting The scale is ...

Page 10: ...elect RS 232 Baud Rate Protocol This scale is equipped with 4 RS232C output baud rate and 2 transmission protocol settings for user to select a Scale is off b Press and hold TARE then press ON ZERO c Scale displays F 1 d Press TARE four times e Scale displays F 5 f Press MODE to select baud rate of 4800 9600 19200 38400 g Press and hold MODE for 2 seconds to enter transmission protocol setup h Pre...

Page 11: ...ZERO c Scale displays F 1 d Press TARE five times e Scale displays F 7 f Press MODE to select transmission mode trAn1 or trAn2 g When the preferred data transmission mode shows on display press ON ZERO to confirm and restart the scale or TARE for next function setup NOTE trAn1 is to transmit data continuously after reading is stabled trAn2 is to transmit single data only after reading is stabled H...

Page 12: ...everything is removed from platter a Scale is off b Press and hold TARE then press ON ZERO c Scale displays F 1 d Press TARE six times e Scale displays F 9 f Press MODE to select between Tr _on or Tr oFF To enable Auto Tare Function press TARE when Tr _on appears To disable Auto Tare Function press TARE when Tr oFF appears g Press ON ZERO to confirm and restart the scale or TARE for next function ...

Page 13: ...y to zero Once ON ZERO is pressed this command will be memorized and activated as soon as the weight detected is stable 5 2 WEIGHING a Before weighing make sure that the ZERO INDICATOR is on If not press ON ZERO b Always place an object onto the platter gently Excessive force applied to platter may cause damages to sensor inside scale c The weight of the object would be displayed automatically d I...

Page 14: ...er one of the following methods Press and hold ON ZERO or Press MODE as described in 5 3 5 4 2 About Smart Power Saving Function This unit is equipped with Smart Power Saving Function When this function is employed the backlight will be turned off when constant weight zero weight is detected and remains stable for about 10 3 seconds Backlight will be automatically turned on again when a weight is ...

Page 15: ...e to zero at anytime when pressing ON ZERO 5 6 TARE OFF THE WEIGHT OF A CONTAINER When a container is used follow the below steps to manually tare off the weight of container and get the net weight a Remove all loads away from platter b Make sure that the ZERO INDICATOR is on c Place the container on the platter d Press TARE After TARE is pressed the NET INDICATOR would appear Refer to 5 2 for wei...

Page 16: ... THE TARE EFFECT a Remove all loads from platter b The tare effect will be cancelled Automatically if automatic tare function is employed or By pressing TARE c After TARE is pressed the NET INDICATOR would disappear 5 8 RECHARGE THE SCALE The scale is equipped with a built in rechargeable battery When the LO BAT INDICATOR appears recharge the scale immediately Failure to do so may cause unrecovera...

Page 17: ... of scale Action Secure both ends of the power adaptor and try again Syndrome After turn on scale displays 00000 and after countdown then power off automatically Check Is the platter is loaded when powered on Action Turn scale off Remove all loads from platter and turn on again Syndrome Scale displays 00000 but cannot obtain zero weight after power on countdown Check Is the scale after by airflow ...

Page 18: ...Turn the scale off Remove all loads from platter and turn on again Check Is there anything obstructing the platter Action Remove all obstacles Syndrome Scale turned off automatically Check Is Auto Power Off Function employed Action Refer to 4 2 2 to disable Auto Power Off Function Check Is the LO BAT INDICATOR on Action Apply the power adaptor Syndrome Weighing result is not accurate Check Is the ...

Page 19: ...l disturbances Check Is the scale calibrated correctly Action Contact your dealer Check Is the correct weight unit employed Action Refer to 5 3 for weight unit setting and usage Syndrome Blank weigh result is display Check Is the scale overloaded Action Remove part of the loads on platter and try again ...

Page 20: ...cally Re calibrate the scale if necessary 4 Note In some countries calibration is restricted to be done by an authorized qualified agent only Contact your dealer for more information 5 It is a good practice to apply the dust cover when operating the scale 6 The scale must be placed horizontally during transportation or long time storage 7 Remove platter from scale before transportation or long tim...

Page 21: ...e following EC directives correspond au modèle décrit dans le certificat d approbation CE de type aux exigences de la directive 90 384 CEE modifiée et aux exigences des directives CE suivantes Diese Erklärung gilt nur in Verbindung mit einer Konformitätsbescheinigung einer benanten Stelle This Declaration of conformity is only valid with a certificate of conformity issued by a notified body Cette ...

Reviews: