PolosSpin150i/200i
rev.: 1.0
Operating instruction
SPS-Europe B.V.
page
25
JH131115
6
Installation and commissioning
DANGER!
Please read completely the manual, containing required safety instructions,
carefully prior to installation, starting and operation of the unit.
DANGER!
Installation and initiation of the electrical equipment have to be effected by
competent and qualified personnel only. All valid national instructions (i.e. VDE)
have to be observed. Prior to all works assure the disconnection of all components
to be installed of power supply.
DANGER!
Ejection of parts (rotating elements)
Chucks designed for vacuum operation may never be run without vacuum. When
operating the system without vacuum, special spin chucks have to be used and
maximum speed has to be reduced.
WARNING!
Important:
Test the values of parameter in a test environment without fluids to
make sure that program and system parameters are correct.
Otherwise you could damage the system.
CAUTION!
To get experience with the system, it is
STRONLGY
recommended to test the unit
with DI water first ONLY before use it with chemicals.
6.1 Safety information
This is a highly sensitive electromechanical appliance. Be careful while unpacking,
installing, starting and operating the unit.
CAUTION!
Heavy weights
Parts included could be heavy and have to be handled properly.
CAUTION!
Before moving the system keep it for at least 24 hours under approved operating
condition to make the unit adapting to ambient temperature. Due to alteration to
temperature during transport unwanted length modification may occur.