PolosSpin150i/200i
rev.: 1.0
Operating instruction
SPS-Europe B.V.
page
61
JH131115
11.3 Facility requirements
Operating voltage
Voltage: 100 -120 VAC or 200 - 240 VAC (auto select)
Frequency: 50 / 60 Hz
Max. current: 2,5 / 5 A
Power consumption (max.): 500 Watt
Motor purge
Required supply: 20-50kPa or 2 - 5l/min
Connector: ID 4mm
– OD 6mm
When using liquids (especially chemicals) the connection (Motor purge) must be
continuously supplied by gas (for this feature is possible to use compressed air,
nitrogen or any other neither flammable nor corrosive gas). This maintains an
overpressure in the drive shaft bearing to protect against liquid ingress.
Note:
This is essential to ensure the long term lifetime and performance of the
system.
Vacuum
Minimum vacuum requirement: - 70 kPa (-21 inchHg). (relative to atmospheric
pressure)
Connector: ID 6mm
– OD 8mm
Minimum supply line diameter: ID 6 mm (¼"), OD 8mm
For the correct operation of vacuum chucks and adapters, the vacuum supply
requirements are:
Drain
Connector: 1" M-NPT external thread
The drain comes supplied with a removable cap.
The drain outlet must lead to a chemically compatible drain or drain store.
An adapter for exhaust separation is optional available.
Programmable relay-contacts
Connector: DSub9 female
Programmable auxiliary potential free relay output suitable for external switching.
Contact Ratings:0.5 A/125 VAC
– 0.3A/60VDC
The Spin 150 provided a cable with DSub9 female connector.
Pin 1 and 2: Programmable relay contact 1 (normally open).
Pin 3 and 4: Programmable relay contact 2 (normally open).
Pin 5 and 6: Programmable relay contact 3 (normally open).
Pin 7: 24 VDC
Pin 8: Ground
Pin 9: not connected.
COMPRESSED AIR/ PNEUMATIC (Optional)
Compressed air is necessary to supply control valves, pneumatic lid or other option
available.
Connector: ID 6mm
– OD 8mm
Requirement: min. 300
– max. 400kPa. Min. 3.0bar – max. 4.0bar.