ECG and Heart Rate Monitoring
élance
Vital Signs Monitor Operations Manual
Page 7-9
3.
Touch
the
required selection
from the list.
4.
Touch
OK
to confirm.
7.9 et
HR
Alarm
S
TEPS
:
1.
Touch
the
ECG Numerical Area
.
A Setup ECG window will appear.
2.
Touch
the
Alarms On
S
You can set the alarm activation and alarm limits from
élance
Vital Signs Monitor.
key
.
Only Yes will be shown, HR alarm cannot be
turned OFF.
T
3.
ouch
the
required Alarm
Selection.
4.
Touch
the
Low Limit
k
5.
Touch
the
High Limit
eys
to increase or decrease the value.
keys
to increase or decrease the value.
6.
Touch
the
OK key
to confirm.
.10 Time to ECG Alarm
7
Alarms
Time to Alarm
ECG - As
seconds
ystole
Less than 7
ECG – Ventricular Fibrillation
Less than 8 seconds
Heart Rate - High Limit Violation
Less than 9 seconds
Heart Rate - Low Limit Violation
Less than 10 seconds