ECG and Heart Rate Monitoring
élance
Vital Signs Monitor Operations Manual
Page 7-8
Notes:
The HR source will be displayed above the hear rate number if the HR source is not
from ECG.
If you use ART as an alternative HR Source, Spacelabs Healthcare recommends
setting up each monitor with only one arterial pressure channel.
Use of SpO
2
as the primary heart rate source is not recommended because SpO
2
is a
7.7 Set Pulse Tone Source
you to activate or deactivate an audible signal for the
élance
generated from the ECG or SpO
2
.
frequent source of false alarm.
The pulse tone enables
ital Signs Monitor. The audible signal can be
V
S
TEPS
:
1.
Touch
the
ECG Numerical Area
.
A Setup ECG window will appear.
.
Touch
the
Pulse Tone Source
2
key
.
A dropdown list, as shown below, will
ear.
app
3.
Touch
the
required Pulse Tone
Selection.
4.
Touch
the
OK key
to confirm.
7.8
E
th
lection
Frequency range accepted
Set ECG Filter
CG filter options provide choice for ECG frequency acceptance. The selections and
eir respective frequency acceptance range are as listed below.
Filter se
Monitor 0.6-
40Hz
Extended 0.05-100Hz
Filter 0.5-30Hz
S
TEPS
:
1.
Touch
the
ECG Numerical Area
.
A Setup ECG window will appear.
2.
Touch the ECG Filter
key. A dropdown list, shown as below, will appear.