Noninvasive Blood Pressure (NIBP) Monitoring
élance
Vital Signs Monitor Operations Manual
Page 14-6
3.
Touch
the
required Format
Selection.
4.
Touch
the
OK key
to confirm.
14.6 Set Initial Inflation Pressure
The initial inflation pressure can be set from 100 to 270 mmHg. The default value is
w to set the initial inflation pressure.
S
TEPS
:
.
Touch
the
NIBP Numerical Area
.
A Setup NIBP window will appear.
h
the
Initial Inflation Pressure
set to 160 mmHg. Follow the steps belo
1
2.
Touc
keys
to increase or decrease the
14.7
set the Systolic alarm activation and alarm limits.
.
Touch
the
NIBP Numerical Area
.
A Setup NIBP window will appear.
value.
3.
Touch
the
OK key
to confirm.
Set Systolic Alarm
ou can
Y
S
TEPS
:
1
2.
Touch
the
Systolic (NIBPs) Alarms On
key
.
A dropdown list, as shown
below, will appear.
3.
Touch
the
requir
n.
4.
Touch
the
Low Limit
ed Alarm
Selectio
keys
to increase or decrease the value.
5.
Touch
the
High Limit
keys
to increase or decrease the value.
6.
Touch
the
OK key
to confirm.