Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-21
Setting Name
Default Value
Possible Values
PR low alarm limit
50
30-235 BPM
NIBPs high alarm limit
Hg
150 mmHg
35-250 mm
NIBPs low alarm limit
mHg
30-245 mmHg
100 m
NIBPs alarms on
Off
On, Off
NIBPd high alarm limit
100 mmHg
15-210 mmHg
NIBPd low alarm limit
60 mmHg
10-205 mmHg
NIBPd alarms on
Off
On, Off
NIBPm high alarm limit
110 mmHg
25-230 mmHg
NIBPm low alarm limit
90 mmHg
20-225 mmHg
NIBPm alarms on
Off
On, Off
NIBP initial inflation pressure 160 mmHg
100 - 270 mmHg in steps of 10 mmHg
NIBP Auto Mode
OFF
4,5, 10, 15, 30, 60
20,240
minutes
OFF, 3,
, 90,1
NIBP Display Format
Sys/Dia
Sys/Dia, Mean
P1s/P1 ARTs/P1 PRSs low
alarm limit
100 mmHg
-10 to 185 mmHg
P1s/P1 ARTs/P1 PRSs high
180 mmHg
190 to 250 m
alarm limit
mHg
P1s/P1 ARTs/P1 PRSs
Off On,
Off
alarms on
P1m/P1 ARTm/P1 PRSm
low alarm limit
70 mmHg
-10 to 130 mmHg
P1m/P1 ARTm/P1 PRSm
high alarm limit
130 mmHg
135 to 230 mmHg
P1m/P1 ARTm/P1 PRSm
Off On,
Off
alarms on
P1d/ P1 ARTd/P1 PRSd low
alarm limit
60 mmHg
-10 to 120 mmHg
P1d/P1 ARTd/P1 PRSd high
mHg
alarm limit
120 mmHg
125 to 210 m
P1d/P1 ARTd/P
alarms on
1 PRSd
Off
Off On,
P1 CVPm/P1 LAPm/ 1
P
it
Hg
RAPm low alarm lim
0 mm
-10 to 225 mmHg
P1 CVPm/P1 LAPm/P1
g
Hg
RAPm high alarm limit
20 mmH
-5 to 230mm
P1 CVPm/P1 LAPm/P
RAPm alarms on
1
Off On,
Off
P2s/P2 ARTs/P2 PRSs low
Hg
mHg
alarm limit
100 mm
-10 to 185 m
P2s/P2 ARTs/P2 PRSs high
mmHg
50 mmHg
alarm limit
180
190 to 2