Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-12
T
v
a very short time. When this function is enabled,
certain parameter limit violations are not considered to be in alarm until they have
existed for the times listed below:
6.7 Alarm Validation
he Alarm Validation function is intended to reduce nuisance alarms in which parameter
alues may exceed the alarm limits for
Limit Violation
Alarm Validation Time
RR High Limit
10 seconds
SpO
2
High Limit
10 seconds
SpO
2
Low Limit
10 seconds
PR High Limit
10 seconds
TEMP High Limit
10 seconds
TEMP Low Limit
10 seconds
:
1.
Touch
the
Monitor Setup
icon from the Main Menu. A Setup menu will appear.
2.
Touch
the
Clinical
from the Setup menu. A Setup clinical window will appear.
3.
Touch
the
Alarms
key. A Setup Alarms window will appear.
4. Touch
the
Alarm Validation
S
TEPS
key.
A dropdown list, as shown below, will
appear.
5.
Touch
the
required selection
to turn alarm validation On or Off.