ECG and Heart Rate Monitoring
élance
Vital Signs Monitor Operations Manual
Page 7-7
3.
Touch
the
uired Size
Selection
4.
Touch
the
OK key
to confirm.
7.6
Set
HR
Source
You can select heart rate sourc
meters,
2
.
vides the capability of auto-switching to alternate
rimary source becomes unavailable. The priority of
ate HR source is as follow:
r
primary source is not ECG, the auto HR source switch
ds after the primary source is unavailable. If all of the alternate
e
ote:
If ART is used as an alternate HR source, it is recommended that the arterial
ressure channel is set up with P1 ART and P2 ART label.
req
.
e from non-ECG para
namely ART and SpO
The
élance
Vital Signs Monitor pro
R source whenever HR from the p
H
the altern
ECG
P1 ART
P2 ART
SpO
2
If the primary source is ECG, 15s after HR artifact or HR lead off message where HR
value becomes unavailable, the HR source will switch to the next available paramete
ith the highest priority. If the
w
will happen 5 secon
sources are unavailable, or if the primary source becomes available, the HR sourc
will be switched to primary source immediately.
For ECG, unavailable means the monitor gets HR artifact or HR lead off alarms. For
other parameters, unavailable means the HR value is blank.
N
p
S
TEPS
:
1.
Touch
the
ECG Numerical Area
.
A Setup ECG window will appear.
2.
Touch
the
HR Source
key
.
A dropdown list, as shown below, will appear.
3.
Touch
the
required HR Source
Selection.
4.
Touch
the
OK key
to confirm.