![Spacelabs Healthcare 93300 Operation Manual Download Page 71](http://html1.mh-extra.com/html/spacelabs-healthcare/93300/93300_operation-manual_1339019071.webp)
Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-19
TEMP Conditions:
Condition
Display Value
Alarm Message
Priority
TEMP within limits
<Number>
None
None
TEMP < Low Limit
<Number>
TEMP lower than Low Limit
Medium
TEMP > High Limit
<Number>
TEMP higher than High Limit
Medium
Temp > 50 degrees C
out of range
+++
TEMP
Low
Temp < 5 degrees C
<Blank>
TEMP unplugged
Low
*TEMP probe
disconnected
--- TEMP
unplugged Low
TEMP bad calibration
resistor
--- TEMP
needs
service
Low
IBP Conditi
ons:
Condition
Display Value
Alarm Message
Priority
IBPs within limits
<Number>
None
None
IBPs < Low Limit
<Number>
IBPs lower than Low Limit
Medium
IBPs > High Limit
<Number>
IBPs higher than High Limit
Medium
IBPm within limits
<Number>
None
None
IBPm < Low Limit
<Number>
w Limit
Medium
IBPm lower than Lo
IBPm > High Limit
<Number>
IBPm higher than High Limit
Medium
IBPd within limits
<Number>
None
None
IBPd < Low Limit
<Number>
IBPd lower than Low Limit
Medium
IBPd > High Limit
<Number>
IBPd higher than High Limit
Medium
No IBP cable connected
(at startup/standby)
<Blank> None
None
*IBP cable disconnected
(during monitoring)
<Blank> IBP
unplugged Low