Vital Signs Monitor Operations Manual
Page 16-4
When a strip chart recording is in progress, you can cancel it by pressing the Record
16.4 T
port title and the date/time of the trend
report.
A “Patient Demographic” trailer is printed. This provides a space for the user to
write the patient’s name and id, as well as comments. If patient has been
admitted, the patient’s name is printed.
16.5 Recorder Messages
larm tone sounds and the message
ecorder needs service” is displayed in the bottom display message area.
When the "Recorder needs service" message is displayed, the cause of the failure is
logged in the event log for viewing by service personnel.
icon.
rend Recordings
When the recorder option is installed, the recorder can print out a Trend Report. This
report contains the same information that is presented in the trends display, with the
following additions:
A header that includes a Trend Re
The date is displayed along with the time of trend data entries.
All trend data entries are printed (including the data stored of up to 120 hours).
Temperature is printed in the currently active units of measure along with the
units of measure.
If the recorder door is open, a technical a
“Recorder door open” is displayed in the bottom display message area.
If the recorder is out of paper, a technical alarm tone sounds and the message
“Recorder out of paper” is displayed in the bottom display message area.
When the recorder is not physically connected, and “Record” icon is touched, a
message “Recorder not connected” will be displayed at the bottom display message
area.
If the monitor senses a recorder error condition, a technical alarm tone sounds and the
message “R