Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-17
pO
2
Conditions:
S
Condition
Display Value
Alarm Message
Priority
SpO , within
2
limits
<Number>
None
None
SpO < Low Limit
2
2
lower than Low Limit
Medium
<Number>
SpO
SpO > High Limit
2
2
<Number>
SpO higher than High Limit
Medium
Bad SpO Probe
2
2
---
SpO replace sensor
Low
Cannot regulate LED
intensity
(after startup/standby)
<Blank> None
None
Cannot reg
intensity
(after finger in probe)
ulate LED
eck sensor placeme
Low
--- SpO
2
ch
nt
Pulsations Too Weak
---
eak signal
Low
SpO
2
w
Probe unplugged
(after startup/standby)
<Blank> None
None
*Probe unplugged
(after finger in probe)
<Blank> SpO
2
unplugged Low
Motion artifact
---
SpO
2
artifact Low
PR within limits
<Number>
None
None
PR
< Low Limit
<Number>
PR lower than Low Limit
Medium
PR > High Limit
<Number>
PR higher than High Limit
Medium
Perfusion too low
---
Low perfusion
(Masimo SpO
2
only)
Low
Poor signal quality
---
Low signal IQ
(Masimo SpO
2
only)
Low
No pulse
---
Pulse Lost
(Nellcor SpO
2
only)
Medium