Printing
élance
Vital Signs Monitor Operations Manual
Page 16-3
ngs
rid has major divisions at 5
m intervals and minor divisions at 1 mm intervals, in both the x and y direction.
The waveform area contains one or two user-selected waveforms. If one waveform is
selected, the waveform uses the entire grid. If two waveforms are selected, each
waveform uses half of the grid, with no overlap. If the same waveform is selected for
both waveform channels, it is treated as a single waveform selection (i.e., one
waveform using the entire grid).
Any displayed waveform may be selected for printing. The printed waveform scaling
and lead-off handling matches that of the displayed waveform. The descriptive text
above the waveform area describes the top waveform. The text below the waveform
area describes the bottom waveform (or is left blank if only 1 waveform is printed).
16.3 Strip Chart Recordi
Strip chart recordings contain a waveform area that includes a 40 mm-wide grid with
descriptive text above and below the grid. The waveform g
m
The descriptive text includes the waveform label, the waveform scale and the
associated
rea includ
parameter value. Additionally, the descriptive text above the waveform
es the current date, printing start time, the waveform delay, and the sweep
s a trailer with the current date, time and a snapshot
railer may be different from those in the descriptive text
they have been taken at the end of the strip chart
a
speed (fixed at 25 mm/sec) and the parameter value at the printing start time.
he strip chart recording contain
T
of all parameters (label, value and units of measurement) that are currently on display.
The parameter values in the t
t the start of the recording because
a
recording.
Alarm-generated strip chart recordings contain a header with an “Alarm Snapshot”
banner and a list of the parameters (label, value and units of measure) that are
currently in alarm.