![Spacelabs Healthcare 93300 Operation Manual Download Page 126](http://html1.mh-extra.com/html/spacelabs-healthcare/93300/93300_operation-manual_1339019126.webp)
Invasive Blood Pressure (IBP) Monitoring
élance
Vital Signs Monitor Operations Manual
Page 13-3
A Setup IBP window, as shown below, will appear.
2.
Set
the required
parameters
.
3.
Touch
the
OK key
to confirm.
13.4 Zero Set
You must zero the system before you can begin monitoring. Zeroing will:
Establish atmospheric pressure at zero
Compensate for hydrostatic effects of any fluids in the catheter-tubing system
S
TEPS
:
1.
Touch
the
IBP Numerical Area
.
A Setup IBP window will appear.
.
Touch
the
Zero Set key
.
A “P1 zeroing” message, as shown below, will appear at
the bottom of the screen.
2