Chapter 6: Safety
The system is intended for use in the electromagnetic environment specified below.
Table 2: Manufacturer’s Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment
Electrostatic
Discharge (ESD)
IEC 61000-4-2
2.0KV, 4.0KV, 6.0KV
contact
2.0KV, 4.0KV, 8.0KV air
2.0KV, 4.0KV, 6.0KV
contact
2.0KV, 4.0KV, 8.0KV
air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
Transient burst
IEC 61000-4-4
2KV on the mains
1KV on signal lines
2KV on the mains
1KV on signal lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
0.5KV, 1.0KV, 2.0KV on
AC power lines to
ground
0.5KV, 1.0KV on AC
power lines to lines
0.5KV, 1.0KV, 2.0KV
on AC power lines
to ground
0.5KV, 1.0KV on AC
power lines to
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
>5% U
T
(>95% dip in U
T
)
for
0.5 cycle
40% U
T
(60% dip In U
T
)
for 5
cycles
70
% U
T
(30% dip in U
T
) for 25
cycles
>5% U
T
(>95% dip In U
T
)
for 5s
>5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip In U
T
)
for
5 cycles
70
% U
T
(30% dip in U
T
) for
25 cycles
>5% U
T
(>95% dip In U
T
)
for 5s
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the SonoSite
ultrasound system requires continued
operation during power mains
interruptions, it is recommended that
the SonoSite ultrasound system be
powered from an uninterruptible power
supply or a battery.
Power
Frequency
Magnetic Field
IEC 61000-4-8
3 A/m
3 A/m
If image distortion occurs, it may be
necessary to position the SonoSite
ultrasound system further from sources
of power frequency magnetic fields or
to install magnetic shielding. The power
frequency magnetic field should be
measured in the Intended installation
location to assure that it is sufficiently
low.
Summary of Contents for titan
Page 1: ...TITAN Ultrasound System User Guide Supplement ...
Page 2: ......
Page 4: ......
Page 5: ...简体中文 TITAN 超声仪用户指南 补充说明 ...
Page 18: ...12 Chapter 6 Safety ...
Page 19: ...Ergänzung zum TITAN Ultraschallsystem Benutzerhandbuch ...
Page 35: ...Suplemento del Manual para el usuario del sistema de ecografía TITAN ...
Page 51: ...Supplément au Guide d utilisation de l échographe TITAN ...
Page 66: ...14 Chapitre 6 Sécurité ...
Page 67: ...Supplemento al Manuale dell utente del sistema per ecografia TITAN ...
Page 82: ...14 Capitolo 6 Sicurezza ...
Page 83: ...Suplemento do Manual do Usuário do Sistema de Ultra som TITAN ...
Page 98: ...14 Capítulo 6 Segurança ...
Page 99: ...Tillæg til brugervejledning til TITAN ultralydssystem ...
Page 113: ...Tillägg till användarhandbok för ultraljudssystemet TITAN ...
Page 128: ...14 Kapitel 6 Säkerhet ...
Page 129: ...TITAN 超声仪用户指南 补充说明 ...
Page 142: ...12 第 6 章 安全性 ...
Page 143: ......
Page 144: ...P06816 01 ...