Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
2 — 8
Diagnostic mode
Diagnostic displays
Software
1.
The software display shows the version of software installed in the pump. It is similar to
the following display:
SOFTWARE VERSION=
2.30
To calibrate the transducer, both a syringe extension set and a mercurial sphygmomanometer
(or a similar pressure reading device) are used.
Insert the pressure sensing disc into the pump’s sensor housing and attach the disc to the
pressure reading device.
In the Diagnostic mode, scroll through the displays until the following display appears:
Calibrate
2.
transducer
PRESS START TO
CALIBRATE TRANSDUCER
Press the
START
button and the following display is shown:
Set the pressure reading device to zero mmHg and then press the
START
button.
The following two displays automatically appears:
ZEROING
PLEASE WAIT
SET ZERO mmHg
AND PRESS START
SET 300 mmHg
AND PRESS START
Set the pressure reading device to 300 mmHg and press the
START
button.
The following three displays automatically appear:
CALIBRATING
PLEASE WAIT
CALIBRATION
COMPLETE
PRESS START TO
CALIBRATE TRANSDUCER
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...