Configuration mode
Available Configuration mode parameters and settings
The syringe parameter allows a choice of any one of eight different syringe brands to
be pre-selected and used during an infusion.
Choice:
BD PLASTIPAK
INJECTOMAT
IMS PUMPJET 30
MONOJECT
BRAUN OMNIFIX
TERUMO
BD PLASTIPAK (A), see page 1-4.
BRAUN PERFUSOR, optional syringe. See Chapter 8.
This parameter allows the five syringe sizes that are available to be locked to any one
size, thus excluding the other four settings, or NOT LOCKED in order to accept all five choices.
Choice:
5 ml ONLY
10 ml ONLY
20 ml ONLY
30 ml ONLY
50/60 ml ONLY
NOT LOCKED
Syringe brands
1.
Syringe size
2.
After the required infusion program values have been set into the pump (i.e. during the
set-up mode) this parameter allows these set-up program infusion values to be tempo-
rarily locked in. This prevents tampering and is useful for home treatment.
The lock must be turned off (i.e. the lock program value set to NO) to change the set-up
infusion mode.
Choice:
YES or NO
Lock program
values
3.
This parameter allows the maximum infusion rate to be set. Above the set rate, an
alarm message is displayed; the pump will not infuse.
The infusion rate that is initially entered into the pump in the set-up mode acts
independently of the maximum infusion rate that can be entered in the configuration
mode.
The configuration mode maximum rate setting takes priority with respect to the
maximum infusion rate that can be used.
The Instruction Manual details the maximum infusion rates for the five different syringe
sizes that are available.
Choice:
10 to 200.0 ml/h in 0.1 ml steps.
Max infusion
rate
4.
(contd.)
WARNING
The pump must be set to operate with the brand and size of syringe that
is going to be used. Using a different brand to that selected could lead
to the incorrect amount of drug being administered, resulting in injury or
death.
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...