Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
5 — 10
Batteries.
Checks and
replacement
Repair procedures
Front and/or
rear case
repair
In March 1999, a new strengthened and modified front and rear case moulding for the 3000
range of syringe pumps (see
Figure 5.2) was introduced.
RIBS ADDED,
EXTRA STRENGTHENING
CORNERS ROUNDED TO MINIMISE
THE DAMAGE CAUSED BY IMPACT
CORNERS ROUNDED
TO MINIMISE THE
DAMAGE CAUSED
BY IMPACT
PILLAR HEIGHT INCREASED
TO MINIMISE STRESS
GM0
9
8
1-A
CORNERS
ROUNDED
TO MINIMISE
IMPACT
GM0
9
8
3-A
If an old style front or rear case becomes damaged and requires replacing then the appropriate
repair kit is available from Smiths Medical. There are two kits (front or rear case) which each
contain all the necessary parts to carry out a repair.
The contents of the repair kits are shown in
Tables 5.3, 5.4
and
5.5.
WARNING
The internal pump batteries must be disposed of in accordance with the
manufacturers instructions. Lead acid batteries must
not be placed in the
normal waste stream.
Smiths Medical recommend that the condition of the three internal batteries is checked at
least annually.
The batteries will normally last several years, but if they should fail to charge then all three
batteries must be replaced at the same time. The batteries are held in place in the front
casing by three-pronged flexible plastic mouldings.
Switch the pump on and fully charge the batteries for at least 14 hours. Remove the AC
power and run the pump at 99.9 ml/hr. If the
LOW BATTERY
alarm appears on the pump's
display before 2.5 hours has elapsed then all three batteries should be replaced, as detailed
below:
1.
Open the case (see page 2-11).
2.
Noting their orientation, prise out the three faulty batteries. Also noting the
connections remove all six spade tags.
3.
Reconnect three new fully charged 2 V, 2.5 AH, lead acid, D Cell batteries by
reversing the steps detailed in (2) above, ensuring that the two rubber packing
spacers that are attached to the pillars are still in place.
4.
Close the casing (see page 2-11).
Checks
Replacement
Figure 5.2 Strengthened front and rear case mouldings
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...