Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
4 — 8
Setting RV1
The procedure for setting potentiometer RV1 is as follows:
1.
Switch off the external AC power, remove the pump’s power connector and open the
casing (see page 2-11).
2.
Remove plug
PL11 and PL12 from the Regulator board and connect a 68 ohm, 1 watt
resistive load across pins 2 and 3 of plug PL11. Pins 2 and 3 are the two middle
pins on the connector.
Note:
A Molex connector (
part number 0053-0658
) enables the 68 ohm resistor to be
easily connected to the pins of PL11.
Setting RV1
WARNING
Only a qualified technician may carry out the following procedures. With the
case open, dangerous voltages are present when AC power is applied.
3.
Connect and switch
ON
the AC power. If necessary, adjust RV1 to give a DC voltage
across the 68 ohm load, dependent on the temperature, as shown in Table 4.3.
RV1 is located near the bottom left hand corner of the board (see
Figure 7.8
).
Table 4.3 Temperature/voltage range for setting RV1
Ambient
Voltage across
temp. (deg. C)
load (volts)
28
7.015
27
7.027
26
7.038
25
7.050
24
7.062
23
7.073
22
7.085
21
7.096
20
7.108
19
7.120
18
7.131
17
7.143
16
7.154
15
7.166
4.
Switch off the AC power. Remove the 68 ohm load and reconnect PL11 and PL12 to the
Regulator board. Assemble the casing (see page 2-11).
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...