Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
2 — 2
Configuration mode
With the pump switched on
and in the set-up mode, complete the following procedures to
enter the Configuration mode.
Press and hold down the
PURGE
button. Press either the
▲
or
▼
button. The following is
displayed:
Within six seconds of pressing the buttons, press the
START
button to confirm the
configuration mode.
Note:
If the
START
button is not pressed within six seconds, the pump returns to
its set-up mode.
A display similar to the following will be shown:
Press the
▲
or
▼
button sequentially to scroll through the settings available within each
parameter. The setting required must be displayed.
The various settings are retained by the pump until the Configuration mode is re-entered
and the setting reset.
Changing a
setting
Press the
ENTER
button to move to the next parameter. If a new choice is required,
repeat the above. Sequentially pressing the
ENTER
button scrolls through the available
Configuration parameters.
To exit the Configuration mode and return to the set-up mode press the
STOP
button at
any time.
Entry into the Configuration mode
SYRINGE TYPE=
BD PLASTIPAK
CONFIGURATION MODE ?
(USE WITH CARE)
Moving to the next
parameter
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...