Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
5 — 4
Cleaning and changing fuses
WARNING
Do not immerse the pump in any liquids. Immediately wipe off any liquid
that may be spilt on the pump.
The outer surfaces of the pump can be cleaned by wiping them over with a damp cloth
(soapy if necessary).
Cleaning
WARNINGS
The following repairs and procedures must only be carried out by qualified
technicians.
Disconnect the pump from the AC supply before opening the casing.
The safety and reliability of the pump may be compromised by the use of parts
other than those specified in this Manual.
•
FS1 is a fuse located in a DC line and has a 1 amp rating.
•
FS2 acts as an AC (time delay) fuse that is rated at 500 mA.
•
FS3 is a fuse located in a DC line and has a 2 amp rating.
Renewal of
fuses
1.
Disassemble the casing (see page 2-11). This will give access to most of the
components on the regulator board.
2.
To renew FS2, the protective covering over the faulty fuse must first be prised off.
The fuse can be removed and replaced with a fuse of the correct ratingand size.
3.
Refit the protective covering.
4.
To renew a faulty FS1 or FS3 (both are situated on the underside of the board), the
three screws that retain the board must be removed.
5.
Carefully lift up the board and turn it over to reveal both fuses. Remove the faulty
fuse and replace it with a fuse of the correct rating and size.
6.
Refit the board and secure with three retaining screws.
7.
Reassemble the casing, being careful not to trap any leads.
Access to
fuses
Repairs
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...