Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
2 — 5
Configuration mode
This parameter allows a selection of in-line pressure sensing options.
If
YES
is chosen, the pump senses the presence of a syringe extension set and shows
the in-line pressure on the display.
If the extension set is not being used, choosing
YES
causes the following prompt to
appear when the
START
button is pressed...
Press
START
to run without disposable
This display serves as a warning that the extension set is not being used. Pressing the
START
button a second time runs the pump without the extension set.
Choosing
NO
also allows the pump to run without the extension set.
Choice:
YES or NO
Use pressure
16.
transducer
This parameter allows the pressure bar graph to be displayed when
YES
is chosen.
If
NO
is chosen, the display will be similar to the following:
10 mmHg (75)
Choice:
YES or NO
Display
17.
pressure
bar graph
If
YES
is chosen this parameter allows the
PURGE
button to infuse a bolus when the
pump is running.
Choice:
YES or NO
Allow bolus
18.
while running
If
YES
is chosen, this parameter allows the infusion rate to be changed during the actual
infusion.
The infusion can thus be titrated to the physiological response of the patient.
Choice:
YES or NO
Allow rate
19.
change while
running
In the Intermittent mode, this parameter allows a start delay between 0 and 24 hours to
be chosen before the intermittent regime is started.
Choice:
YES or NO
Intermittent
20.
mode
start
delay
This parameter allows a choice of Baud rates when the pump is connected to a PC or a
printer.
Choice:
9600
4800
2400
1200
600
300
RS232 Baud
21.
rate
(contd.)
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...