2 — 13
Occlusion thrust
Thrust checks
Figure 2.2 Thrust measuring set up
The following thrust checks use the weights that correspond to the factory-set occlusion
threshold levels for a 3200 (i.e. 3.5 and 4.2 kg). If a different occlusion level setting is required,
the weights must be adjusted accordingly.
The thrust adjustment procedures are shown on page 2-14.
1. Set the pump's plunger clamp to approximately midway along its support tube.
2. Remove the plunger from a BD 60 ml syringe and saw the end off the syringe. Place the
modified syringe onto the pump so that it will act as a guide for the weight support rod and
also position the size sensor flag (see
Figure 2.2).
3. With the pump switched
ON
, set the infusion rate to 200 ml/hr.
4. Place the pump in a vertical position, with its left hand side uppermost (see
Figure 2.2).
5. Place the weight support rod through the modified syringe and onto the pump's plunger
clamp.
6. Place a weight of 3.5 kg on top of the weight support and check that the pump operates for
between 30 and 60 seconds and does
not
occlude (i.e. the alarm does not sound). Remove
the weight.
7. Place a weight of 4.2 kg on top of the weight support and check that within 60 seconds the
pump does occlude (i.e. the alarm sounds).
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...