Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
Appendix, page A - 5
Appendix, SSF
Figure A.3 Size Sensor Flag: general details
The Size Sensor Gauge set (
part number 0131-0202) contains the gauges that are used to
carry out the Size Sensor tests. These gauges are listed in
Table A.1.
Table A.1
3200 Syringe Size Sensor Gauges (white)
PART No.
GAUGE DIA. (mm) CORRECT RESULT
0131-0165
11.20
No LED lit
0131-0166
12.50 - min.
5ml
LED lights
0131-0167
15.00 - max.
5ml
LED lights
0131-0168
16.20 - min.
10ml
LED lights
0131-0169
18.20 - max.
10ml
LED lights
0131-0170
21.00 - min.
20ml
LED lights
0131-0171
23.00 - max.
20ml
LED lights
0131-0172
23.70 - min.
30ml
LED lights
0131-0173
26.80 - max.
30ml
LED lights
0131-0174
29.00 - min.
50/60ml
LED lights
0131-0175
33.10 - max.
50/60ml
LED lights
Apply a small amount
of loctite 414 to the
security cap
Apply a small amount
of loctite 7400 to the
thread of the grub screw
prior to setting
GM1111-A
X
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...