Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
Appendix, page A - 1
Appendix, SSF
APPENDIX
FITTING a MODIFIED SIZE SENSOR FLAG
In a continual and ongoing programme of improvements to their 3000 range of syringe
pumps, Smiths Medical engineers have introduced a modified Syringe Size Flag (SSF)
1
that allows the size sensor mechanism to be accurately aligned (see
Figure A).
On recently manufactured pumps, a new thicker size sensor opto moulding and two associ-
ated longer fixing screws have been introduced at the same time as the modified SSF, in
order to make the alignment procedures easy.
This Appendix enables a qualified Technician to fit the new SSF; the associated SSF shim/s,
the new thicker size sensor opto moulding (if required), and to complete the final test
procedures.
The SSF and shims etc. required to modify a pump are supplied in kit form (
part number
0137-0025). The spares kit includes:
•
SSF,
part number 0132-0090 (see Figure A.1)
•
0.6 mm thick shim,
part number 0130-0107
•
1.0 mm thick shim,
part number 0130-0108
•
1.2 mm thick shim,
part number 0130-0190
•
1.4 mm thick shim,
part number 0130-0185
•
Opto housing,
part number 0130-0024 (4.65 mm thick)
•
Screw, no. 4 x 5/8 inch, pozi, self-tap,
part number 5017-3410 (2 off)
•
Grub screw, M4 x 6 nylon,
part number 0131-0144
•
Tamper proof protective cap,
part number 0131-0136
•
Instruction leaflet,
part number 0131-0217.
Introduction
Figure A.1 New modified Syringe Size Sensor flag
Part Number
0132-0090
GM1110-A
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...