Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
2 — 1
CHAPTER 2
CONFIGURATION, DIAGNOSTICS and OCCLUSION THRUST
Configuration mode
Configuration mode
The Configuration mode allows the various different pump parameters to be displayed, which
in turn allows various settings within these parameters to be set to the values required for the
infusion.
The full list of Configuration parameters is shown below, but the availability of some parameters
depends upon the setting of other parameters. Each parameter is further detailed in the
sections that follow this list:
1.
Syringe brand.
2.
Syringe size.
3.
Lock program values.
4.
Max infusion rate.
5.
Select pump modes.
6.
Preset volume mode with time.
7.
KVO rate.
8.
Allow mass unit.
9.
Infusion units, mg/kg/h.
10.
Infusion units, µg/kg/h.
11.
Infusion units, mg/kg/min.
12.
Infusion units, µg/kg/min.
13.
Infusion units, mg/h.
14.
Infusion units, µg/h.
15.
Show rate in ml/h while infusing.
16.
Use pressure transducer.
17.
Display pressure bar graph.
18.
Allow bolus while running.
19.
Allow rate change while running.
20.
Intermittent mode start delay.
21.
RS232 Baud rate.
22.
Communication mode.
23.
Pump ID.
24.
Set key beep volume.
25.
Pressure units.
26.
End of menu.
Introduction
WARNING
The Configuration mode must only be used by personnel who are
adequately qualified and have had previous training in the use of the 3200.
Summary of Contents for Graseby 3200
Page 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Page 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Page 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Page 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Page 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Page 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Page 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Page 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...